B2A5 Indicator Display Instruction manuals
Share
B2A5 Meter Calibration Procedure
1. Connect the meter to the correct power supply and link the sensor correctly.
2. Set the parameter through SET and enter SH parameter for loading calibration.
3. Set the SH parameter to fixed parameter, then give the sensor fixed weight, click SET for calibration. (For example, for a 200 kg sensor, usually set the parameter to 50.00, then put a 50 kg fixed weight on the sensor and click SET to set.)
4. Through the Sd parameter, adjust the decimal point of the display unit to the appropriate position.
5. Through the Id parameter, set the index value, both the smallest unit displayed. Large range sensors are recommended to choose the appropriate index value to avoid the meter display figures jumping during use.
Note: The meter needs to be used within the appropriate voltage, do not exceed the rated voltage. The sensor should be used within the range, over range will damage the sensor.
1. Connect the meter to the correct power supply and link the sensor correctly.
2. Set the parameter through SET and enter SH parameter for loading calibration.
3. Set the SH parameter to fixed parameter, then give the sensor fixed weight, click SET for calibration. (For example, for a 200 kg sensor, usually set the parameter to 50.00, then put a 50 kg fixed weight on the sensor and click SET to set.)
4. Through the Sd parameter, adjust the decimal point of the display unit to the appropriate position.
5. Through the Id parameter, set the index value, both the smallest unit displayed. Large range sensors are recommended to choose the appropriate index value to avoid the meter display figures jumping during use.
Note: The meter needs to be used within the appropriate voltage, do not exceed the rated voltage. The sensor should be used within the range, over range will damage the sensor.